MEC2 INVERTED METALLURGICAL MICROSCOPE
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 6
DETAILED ASSEMBLY PROCEDURE
Attaching the Mechanical Stage
And X-Y Controls
The mechanical stage may be installed in either
side of the stage to enlarge the stage surface.
The x-y controls must be installed on the
opposite side of theauxiliary stage.
Generally, the x-y controls are installed on the
right side for comfortable operation.
Installing The Mechanical Stage
Screw the clamping bolts into to the mechanical
stage, then into the fixed stage from below on
either the right or left side. Tighten the bolts with
a screwdriver until the mechanical stage is
securely attached.
Installing The X-Y Controls
Install the ruler using the same method as the
mechanical stage.
Installing and Replacing the Halogen
Bulb (Figures 1-3
Use only the specified halogen lamp: 6V 30W
1. Pull out the lamp house (Figure 1)
2. Holding the bulb with a piece of gauze or
other protective material, insert the bulb pins
straight and fully into the pin holes on the
lamp house. (Figure 2)
CAUTION: The bulb, lamp housing and
nearby parts will be very hot during usage.
3. When replacing the lamp, turn the main
switch to “O (off) and remove the power cord.
Allow the lamp, lamp house and adjacent areas
to sufficiently cool before handling. The lamp
can become very hot and will cause burns.
Fig. 1
Fig. 2
Fig. 3