MEC2 INVERTED METALLURGICAL MICROSCOPE
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 14
Light Path
Selecting Lever
Illumination Application
Pushed In 100% for binocular observation Binocular observation
Pulled Out 20% for binocular observation
80% for video or
microphotography
Binocular observation and video
(CCTV) or microphotography
simultaneously
Polarization Observation
1. Insert the polarizer and analyzer as described on page 8 (Figures 10 & 11).
2. The analyzer hand wheel can be rotated from 0 °to 90°. When the field of view becomes the darkest,
the polarization position has been obtained and polarized observation can be performed.
MICROSCOPE VIDEO
AND PHOTOGRAPHY
Video Microscopy
Selecting the Light Path (Figure 24)
1. Push in the light path selector (1) completely.
2. Using binocular observation, focus your
specimen.
3. Pull out the light path selector completely.
Installing the Video Adapter (Figure 25)
1. Loosen the thumb screw on the trinocular
viewing tube, and remove the dust cap.
2. Remove the dust covers on the both ends of the
video adapter.
3. Insert the flanged end into the trinocular viewing
head and tighten the thumb screw.
4. The threaded video C-mount should be facing
upwards.